UGRA NEC PA241W

Page 1

UGRA Display Analysis & Certification Tool Report Basics Date: Report-Version: Monitor-Name: Profile: Created: Measurement device:

2010-4-24 21:12:39 v1.3.1 \\.\DISPLAY1 C:/Windows/system32/spool/drivers/color/PA241W 02100015UB.icc 2010-4-24 21:00 eye-one display 2

Summary The monitor has passed the certification according to the UGRA DACT specifications. Calibration White Point Gray balance Profile quality

yes yes yes

MultiColor, HighBody Offset/Gravure Paper Type 1/2 Offset on uncoated paper Newspaper Printing sRGB AdobeRGB ECI-RGB

yes yes yes yes yes yes yes

Softproofing

Diagram WhitePoint Gray Balance Profile Quality Softproof Quality

www.microgamma.com


Whitepoint The whitepoint should be as close as possible to the black body curve and the calibration target.The maximum allowed distance to the target whitepoint is DeltaE 2.0. XYZ: XYZ (normalized): Luminance: Next Temperature: Assumed Target Whitepoint: Distance to assumed Target Whitepoint:

117.34 123.94 119.92 94.67 100.00 96.76 123.9 Cd/m2 5837 Kelvin 5800 Kelvin 1.4 deltaE

Blackpoint The blackpoint is not defined in ISO 12646. Therefore UDACT does only measure but not assess it. Luminance: Chromaticity:

0.2 Cd/m2 2.1 Chroma (Lab)

Gray balance Average and maximum calculation will respect measurements with 1% minimum luminance only. The maximum allowed deviations to comply with this test are an average of DeltaC 1.0 and a range of DeltaC 2.0. % 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100

Kelvin 14635 7170 6241 6018 5901 5877 5849 5855 5835 5807 5836 5818 5777 5791 5792 5785 5805 5805 5823 5837 5837

Average Range

5857

Cd/m2 0.21 0.70 2.08 4.20 6.99 10.12 14.26 18.95 24.16 29.52 35.73 42.58 50.04 57.21 65.49 74.37 83.72 92.80 102.83 113.51 123.94

L 1.53 5.12 13.69 21.55 28.47 34.33 40.41 46.03 51.26 55.91 60.63 65.24 69.73 73.65 77.78 81.84 85.78 89.33 93.00 96.65 100.00

Chroma 2.14 2.45 1.48 1.21 0.96 1.21 0.67 0.38 0.51 0.66 0.59 1.14 0.66 0.86 0.94 0.90 0.68 0.64 0.43 0.18 0.00

Gamma

0.74 1.57

1.79

1.84 1.82 1.81 1.80 1.81 1.80 1.80 1.79 1.80 1.80 1.78 1.78 1.79 1.78 1.77 1.75 1.77 1.76 1.70

Tone values = 100.0% Through the calibration of a display tone values can be lost. A display for the printing industry should show at least 95% of the incoming tone values.

www.microgamma.com


Profile Quality This test displays and measures RGB values and compares them with the transformation of the profile. The maximum allowed deviations to comply with this test are an average of DeltaE 3.0 and a maximum of DeltaE 6.0. The assumed chromatic adaptation is: Bradford RGB 0 0 0 0 0 128 0 0 255 0 128 0 0 128 128 0 170 255 0 255 0 0 255 170 0 255 255 85 85 85 128 0 0 128 0 128 128 128 0 128 128 128 128 128 255 128 255 128 170 0 255 170 170 170 170 255 0 170 255 255 255 0 0 255 0 170 255 0 255 255 128 128 255 170 0 255 170 255 255 255 0 255 255 170 255 255 255 170 85 85 85 170 85 85 85 170 85 170 170 170 85 170 170 170 85

Lab 1.5 0.2 -2.1 12.9 49.0 -77.7 26.6 76.7 -119.4 52.7 -84.4 56.5 54.5 -59.2 -10.7 69.4 -46.3 -48.7 87.9 -129.1 87.7 89.1 -108.5 19.7 90.3 -90.3 -15.8 44.3 0.7 -0.5 31.2 75.9 45.7 34.4 83.1 -42.1 59.5 -9.9 66.5 61.1 0.6 -0.5 64.1 21.7 -57.2 91.8 -73.7 43.1 47.1 106.2 -84.6 75.2 0.9 -0.4 93.1 -59.7 95.4 95.3 -36.6 -7.9 55.2 115.0 78.4 57.5 120.4 -26.4 59.8 125.5 -62.9 73.4 62.9 19.5 80.6 33.1 90.0 83.5 48.2 -25.5 98.0 -15.8 102.5 99.0 -7.8 35.1 100.0 0.0 0.0 54.2 50.6 15.4 68.6 -58.2 33.9 46.7 17.5 -45.6 69.9 -43.4 -9.1 56.0 58.2 -30.1 74.1 -8.1 42.6

deltaLab -1.5 -0.2 2.1 -0.8 2.1 -1.3 -0.2 0.5 -0.1 0.0 -1.6 1.8 -0.2 -0.5 0.1 -0.2 0.2 -0.3 0.0 -1.0 0.4 -0.0 -0.7 0.1 -0.0 0.0 -0.2 -0.3 -0.7 0.5 -0.2 0.5 2.9 -0.2 0.1 0.6 -0.1 -0.9 1.0 -0.4 -0.6 0.5 -0.4 -0.0 -0.3 -0.1 -1.4 0.8 -0.1 -0.1 0.1 -0.3 -0.9 0.4 -0.1 -1.4 -0.2 -0.1 -1.2 -0.0 -0.1 0.4 -0.1 -0.0 0.2 0.8 0.0 0.3 0.2 -0.3 0.6 0.7 -0.0 -0.2 -0.2 -0.2 0.6 -0.0 0.1 -0.6 -0.4 -0.0 -0.4 0.7 -0.0 0.0 -0.0 -0.4 -0.6 0.5 -0.2 -0.9 0.7 -0.4 -0.4 0.3 -0.3 -0.6 0.4 -0.5 -0.1 0.3 -0.3 -0.9 0.6

Average Maximum

deltaE 2.6 2.5 0.6 2.4 0.5 0.4 1.1 0.7 0.2 0.9 3.0 0.7 1.3 0.9 0.5 1.6 0.2 1.0 1.5 1.2 0.4 0.8 0.3 0.9 0.3 0.6 0.7 0.8 0.0 0.9 1.1 0.6 0.8 0.5 1.1 1.0 3.0

Gamut-Volume These measurements are only informative. Gamut-Volume ISO sRGB AdobeRGB ECI-RGB v1.0

100 % 100 % 100 % 91 %

www.microgamma.com


ISO-Gamut This test displays and measures Lab values and compares them with the reference. The maximum allowed deviations to comply with this test are an average of DeltaE 4.0 and a minimum Gamut volume of 90% for ISOcoated. Reference 55.0 -37.0 -50.0 66.9 -24.7 -37.1 79.7 -12.5 -21.8 48.0 74.0 -3.0 60.8 50.6 -6.7 76.4 25.8 -6.9 89.0 -5.0 93.0 90.3 -4.7 62.6 92.2 -3.5 31.1 53.1 37.7 28.9 41.5 22.7 16.8 31.9 40.0 24.0 32.5 44.5 -1.8 51.3 1.3 44.5 34.6 -36.4 13.9 36.0 -26.2 -20.9 20.9 9.6 -23.6 89.0 0.0 -1.9 82.8 0.0 -1.7 69.3 0.0 -1.4 54.1 0.0 -1.0 36.6 0.0 -0.5 16.0 0.0 0.0 24.0 22.0 -46.0 40.9 17.9 -36.6 63.7 10.3 -23.8 47.0 68.0 48.0 58.5 47.1 37.9 74.2 22.9 21.4 50.0 -65.0 27.0 62.1 -39.8 21.0 77.0 -19.1 11.0 71.2 18.9 17.3 71.2 22.1 73.1 47.7 71.2 16.2 38.0 55.4 -20.9 73.7 -22.8 67.6 52.3 -52.3 -20.1 43.3 -17.0 -48.6 95.0 0.0 -2.0 88.5 -0.4 -3.1 82.0 -0.9 -4.1 67.7 -2.0 -4.4 52.2 -2.5 -3.5 37.5 -3.9 -3.1 26.3 -6.8 -3.4 Average Gamut-Volume

Lab 55.6 -31.8 -49.3 67.3 -24.8 -36.9 80.0 -11.8 -22.0 48.4 74.0 -3.7 61.2 51.0 -7.2 76.9 26.1 -7.1 89.1 -4.1 92.5 90.6 -4.3 61.9 92.5 -2.7 30.8 53.4 38.7 28.4 42.0 22.9 16.1 32.0 40.2 22.5 32.8 45.0 -2.6 51.5 1.9 43.4 35.0 -35.4 12.9 36.3 -25.6 -20.9 21.6 9.7 -23.8 89.3 0.6 -2.1 83.1 0.6 -2.0 69.6 0.6 -1.1 54.4 0.7 -1.4 36.8 0.9 -1.4 16.8 0.4 -1.2 24.4 22.4 -46.2 41.3 18.3 -37.1 63.9 11.5 -24.3 47.3 68.3 46.5 59.0 47.1 37.3 74.5 23.5 20.8 50.3 -64.2 26.4 62.3 -38.8 20.1 77.2 -17.7 10.9 71.8 19.0 17.2 71.4 23.4 72.5 48.0 71.3 15.3 38.5 55.5 -21.1 73.9 -22.1 66.9 52.6 -50.9 -20.2 43.8 -16.8 -48.6 95.2 0.4 -2.4 88.9 0.0 -3.1 82.3 -0.5 -4.2 68.1 -1.6 -4.1 52.5 -1.8 -3.8 37.7 -3.1 -4.2 26.6 -5.4 -4.8

deltaLab 5.3 0.5 0.9 0.8 0.7 0.6 1.0 0.8 0.9 1.2 0.8 1.5 1.0 1.2 1.4 0.7 0.8 0.7 0.7 0.7 0.9 1.3 1.5 0.6 0.8 1.3 1.6 0.8 0.8 1.1 1.4 1.4 0.6 1.4 1.0 0.5 1.0 1.4 0.5 0.6 0.6 0.5 0.5 0.8 1.3 2.0 1.1 100 %

www.microgamma.com


Measurement Data This table lists all RGB measurements. The XYZ values represent the values from the measurement device. RGB 255 255 255 0 0 0 12 12 12 25 25 25 38 38 38 51 51 51 63 63 63 76 76 76 89 89 89 102 102 102 114 114 114 127 127 127 140 140 140 153 153 153 165 165 165 178 178 178 191 191 191 204 204 204 216 216 216 229 229 229 242 242 242 0 0 128 0 0 255 0 128 0 0 128 128 0 170 255 0 255 0 0 255 170 0 255 255 85 85 85 128 0 0 128 0 128 128 128 0 128 128 128 128 128 255 128 255 128 170 0 255 170 170 170 170 255 0 170 255 255 255 0 0 255 0 170 255 0 255 255 128 128 255 170 0 255 170 255 255 255 0 255 255 170 170 85 85 85 170 85

XYZ 117.34 123.94 119.92 0.21 0.21 0.37 0.68 0.70 0.86 1.98 2.08 2.19 4.02 4.20 4.28 6.67 6.99 6.96 9.70 10.12 10.09 13.58 14.26 13.99 17.99 18.95 18.49 22.97 24.16 23.55 28.12 29.52 28.72 33.97 35.73 34.85 40.65 42.58 41.71 47.62 50.04 48.33 54.52 57.21 55.55 62.43 65.49 63.63 70.86 74.37 72.11 79.62 83.72 81.19 88.22 92.80 89.94 97.60 102.83 99.71 107.56 113.51 110.01 5.73 2.14 31.22 19.50 6.96 108.09 8.47 25.87 3.37 14.15 28.19 34.63 33.40 50.41 112.86 28.90 89.60 10.80 38.27 93.05 62.77 48.18 96.71 118.27 16.65 17.43 17.12 20.11 8.02 0.68 25.88 10.04 32.32 28.69 34.00 3.73 34.59 36.34 35.58 48.40 41.32 111.88 55.05 99.81 43.12 53.12 20.20 109.08 57.46 60.28 58.76 62.80 103.01 11.36 82.29 109.99 119.41 69.15 27.44 1.48 78.53 30.71 54.28 88.45 34.21 109.48 83.54 55.85 36.66 83.16 70.60 6.57 102.74 77.99 115.02 98.08 116.86 11.96 107.59 120.42 64.92 40.83 27.00 17.68 26.54 48.30 20.84

RGB 85 85 170 85 170 170 170 85 170 170 170 85 0 125 207 83 156 220 153 191 232 173 54 102 187 103 142 207 162 190 236 217 14 236 222 90 236 228 159 157 90 63 106 71 56 97 42 30 99 41 63 115 102 35 33 73 45 25 74 97 37 36 67 215 216 220 194 195 199 152 153 155 110 110 112 68 68 69 30 30 30 40 37 107 79 72 138 138 132 183 171 56 28 186 99 62 206 156 127 30 117 56 90 145 94 155 184 153 190 150 126 205 147 26 173 55 72 119 46 104 156 175 39 0 121 140 17 90 167 236 237 242 212 215 221 189 193 201 143 149 156 101 106 111 65 71 75 41 49 52

XYZ 23.40 19.88 54.62 33.31 50.81 58.26 47.50 29.46 55.23 50.71 57.89 21.38 21.51 29.83 77.44 36.47 46.52 88.31 62.09 70.80 99.11 40.60 20.78 22.81 53.00 36.21 41.60 72.85 63.38 70.20 82.67 91.54 10.12 86.46 95.45 26.78 93.03 100.94 56.95 35.65 26.17 11.41 18.79 15.31 8.83 13.94 8.61 3.30 15.54 9.06 9.80 23.02 24.21 5.73 5.83 10.59 6.28 7.81 11.50 20.50 4.96 4.29 10.51 88.10 92.61 92.75 73.53 77.29 77.41 47.47 49.87 49.38 26.43 27.71 27.76 11.22 11.69 11.86 2.68 2.80 2.88 7.64 5.37 23.40 18.18 15.10 37.75 43.07 40.72 63.99 36.54 19.57 3.74 46.77 32.88 11.67 65.66 58.43 37.04 10.02 23.31 10.08 24.70 38.19 22.59 52.88 64.29 50.23 58.26 53.31 36.12 58.64 52.32 7.04 38.83 20.32 12.96 23.13 12.71 22.90 44.56 57.43 10.01 14.33 25.98 39.78 14.27 17.43 52.53 103.85 109.31 109.74 86.98 91.75 93.41 71.40 75.57 78.59 44.17 47.21 49.66 23.80 25.56 27.17 11.22 12.30 13.56 5.37 6.17 7.20


Uniformity This informative-only test shows the luminance (in %) and color deviations (in DeltaC) on the entire screen area based on white, gray and dark measurements in 9 regions. Results of the white measurements: Absolute Luminance 127.27 Cd/m2 Luminance to Centre 102 % Lab-Chroma to Centre 3.17

128.40 Cd/m2 103 % 3.13

128.16 Cd/m2 103 % 2.90

Absolute Luminance Luminance to Centre Lab-Chroma to Centre

127.60 Cd/m2 102 % 3.05

124.77 Cd/m2 100 % 0.00

127.05 Cd/m2 102 % 3.39

Absolute Luminance Luminance to Centre Lab-Chroma to Centre

124.06 Cd/m2 99 % 3.79

124.05 Cd/m2 99 % 3.29

123.70 Cd/m2 99 % 4.44

Results of the gray measurements: Absolute Luminance Luminance to Centre Lab-Chroma to Centre

36.86 Cd/m2 101 % 1.95

37.23 Cd/m2 102 % 1.93

36.79 Cd/m2 101 % 1.74

Absolute Luminance Luminance to Centre Lab-Chroma to Centre

36.71 Cd/m2 101 % 1.81

36.37 Cd/m2 100 % 0.63

36.49 Cd/m2 100 % 1.98

Absolute Luminance Luminance to Centre Lab-Chroma to Centre

35.91 Cd/m2 99 % 1.83

35.74 Cd/m2 98 % 1.78

35.75 Cd/m2 98 % 2.33

Results of the dark measurements: Absolute Luminance Luminance to Centre Lab-Chroma to Centre

10.19 Cd/m2 101 % 1.30

10.29 Cd/m2 102 % 1.47

10.16 Cd/m2 101 % 1.54

Absolute Luminance Luminance to Centre Lab-Chroma to Centre

10.18 Cd/m2 101 % 1.15

10.07 Cd/m2 100 % 1.28

9.98 Cd/m2 99 % 0.86

Absolute Luminance Luminance to Centre Lab-Chroma to Centre

9.93 Cd/m2 99 % 1.26

9.89 Cd/m2 98 % 0.97

9.91 Cd/m2 98 % 1.59

Average 1% 2

Max 3% 4

Summary: Luminance Deviation Chroma Deviation

www.microgamma.com